Electromagnetic Flowmeter - AMF900-P0010-S2N84-NNN-N

 Electromagnetic Flowmeter

SPECIFICATION



Size:Electrode & Grounding:
10A,10,15,20, 25,32,40,50,65,80,100,125Stainless Steel 316L
150,200,250,300,350,400,450,500,600Hastelloy B
700,800,900,1000,1200,1400,1600Hastelloy C
1800,2000 mmTitanium
Measuring Range:Tantalum
Velocity   0 - 0.25 m/s min.Platinum
                0 - 12  m/s max.Tungsten Carbide
Material:Grounding Electrode:
Measuring TubeDN10A/10: use Grounding Ring *1pc
Stainless Steel 304DN15-2000: 3 Electrodes
FlangeProcess Connection:
Carbon Steel (Standard)Flange 
Stainless Steel 304 (Option)Flange Type:
Stainless Steel 316, 316L (Option)JIS 10K / JIS 20K / JIS 40K
Coil HousingANSI 150# / ANSI 300# / ANSI 600#
Carbon Steel (Standard)DIN PN 10 / PN 16 / PN25 / PN 40
Stainless Steel 304 (Option)Grounding Resistance:
Stainless Steel 316,316L (Option)Must be  less than 10 
Liner:Accuracy:
Polyurethane (25-400 mm)+/-0.4% of reading (Velocity >=0.5 m/s)
Neoprene (100-2000 mm)+/-0.0025 m/s (Velocity < 0.5 m/s)
FEP (10A-300 mm)+/-0.2% of reading
PTFE (10A-2000 mm)(Suitable for AMC3200/3200E converter )
PFA (10A-300 mm)Temperature:
Protection Class:-10 ~ +60 oC(Polyurethane, PU)
Junction box IP67, Body IP68-20 ~ +70 oC(Neoprene)
IP 68 (Submersible)-40 ~ +120 oC(PTFE 3/8"to 10")
Conductivity:-40~+140 oC(PTFE 3/8"to6", FEP 3/8"to12")
More than 5 uS/cm-40~+180 oC(PTFE 3/8"to3", PFA3/8"to12")
Explosion Proof:Ambient Temperature:
Exd(ib)ibqIIBT5 -25 to +60 oC
(with AMC3200 only)Max. Pressure:
Cable Entry:350 Kg/cm2
Standard :PG11  Option:1/2"NPTF